The KOBOLD MIS is an economical, full-bore magnetic flow meter that measures, monitors, and batches conductive liquids in 2 to 8 inch flanged pipelines. Also called an electromagnetic flow meter or a magnetic inductive flowmeter, it operates on Faraday's Law, covers flow rates from 2 GPM to 4,400 GPM, and pairs an obstruction-free flow tube with two configurable outputs, standard IO-Link communication, and a rotatable color display.
Designed for water treatment, wastewater processing, filtration systems, and general industrial applications, the MIS combines monitoring, transmitter, and batching functions in a single device. Because measurement is electromagnetic, readings are unaffected by fluid density, viscosity, or temperature variations.
Flow Ranges and Pipe Sizes
Five ANSI Class 150 flange sizes cover 2, 3, 4, 6, and 8 inch pipelines, with measuring spans reaching from 2...277 GPM in the smallest size up to 18...4,400 GPM in the largest. According to specifications, the full-bore design produces a maximum pressure drop of only 0.36 PSI at a flow velocity of 9.8 ft/s. Installation in pipes with larger nominal sizes is also possible using pipe tapers, with the resulting pressure loss taken into consideration.
Accuracy and Measurement Stability
According to specifications, the MIS delivers accuracy of ±(0.5% of reading + 0.5% of full scale) with repeatability of ±0.2% of full scale and a response time under 250 ms. Media viscosities up to 70 cP are within its ratings, the required minimum conductivity is 20 µS/cm, and built-in empty pipe detection recognizes a line that is not completely filled. Minor deposits on the measuring tube generally do not affect measuring accuracy unless their conductivity deviates significantly from the liquid.
Construction and Environmental Ratings
A powder-coated aluminum housing rated IP67 suits washdown environments. The MIS handles media temperatures from 14°F to 158°F, ambient temperatures from 14°F to 140°F, and line pressures up to 232 PSIG. Standard construction includes an NBR liner and Hastelloy C276 electrodes for broad compatibility with conductive process liquids. Installed weights range from 21 lb for the 2 inch meter to 107 lb for the 8 inch version.
U-PACE Electronics, Outputs, and IO-Link
The universal U-PACE electronics provide two arbitrarily configurable outputs, selectable as 4-20 mA, 0-20 mA, 0-10 VDC, or 2-10 VDC analog signals, or as switching, pulse, or frequency outputs. IO-Link V1.1 communication is standard. A 1.4 inch color TFT display rotates in 90-degree increments, refreshes at an adjustable 0.5 to 10 second rate, and is operated through four optical touch keys that work with gloves.
Batching and Bidirectional Totalization
An integrated batching function enables precise dispensing with external PLC control, so simple filling tasks run without a separate batch controller feeding the display's intuitive setup menu. Bidirectional measurement supports lines where flow reverses, and grand and resettable totalizers track volume with the flow rate, daily volume, and total volume shown in the units of measurement the operator selects.
Magnetic Flow Meter Applications
The MIS is designed for water tapping and water treatment, water distribution networks with leakage detection management, wastewater treatment, and filtration systems such as reverse osmosis and ultrafiltration, alongside general industrial applications. It is easy to commission via a clear guided menu and is available with quick delivery. For highly soiled fluids, a bypass pipe with a blind flange lets the line be cleaned without dismounting the flowmeter.
Common Questions about MIS - Economical Full-Bore Magnetic Flow Meter - Flange - for Liquids
Per Faraday's law of magnetic induction, a voltage is induced in a conductor moving through a magnetic field. The conductive liquid acts as that conductor: two sensing electrodes pick up the induced voltage, which is proportional to flow velocity, and volumetric flow is calculated from the pipe's defined cross-section.
A minimum electrical conductivity of 20 µS/cm is required, and the electronics monitor conductivity continuously. If it falls below the minimum, the display shows an Empty pipe error and the flow reading is set to zero. Entrained gas bubbles are measured as volume, so vent the line where necessary.
Measuring ranges by flange size are 2...277 GPM for 2 inch, 2.65...700 GPM for 3 inch, 4.40...1,100 GPM for 4 inch, 12...2,810 GPM for 6 inch, and 18...4,400 GPM for 8 inch, each supplied with ANSI Class 150 flanges. Limited quantities of the 6 and 8 inch sizes are available; consult the factory.
Plan for a straight inlet section of 5 x DN upstream and an outlet section of 3 x DN downstream of the meter. Mounting position is otherwise unrestricted: the MIS installs in all directions, ideally where the pipe always remains completely filled, such as an ascending line.
The meter runs on 19 to 30 VDC with internal power consumption of 200 mA maximum, connected through a single M12x1 4-pin plug. When operated under the CSA label, the external power supply must fulfill IEC/CSA/UL 61010-1 requirements, for example a Class 2 Limited Power Source.
Batching uses the two configurable connections: OUT2 serves as the high-active batching output while OUT1 acts as the control input. A high signal of 0.5 to 4 seconds triggers START or STOP, and one longer than 5 seconds triggers RESET, letting a PLC run simple filling tasks.
The MIS carries cCSAus certification for ordinary locations (ORDLOC, certificate 80190658) and international IECEE certification according to the CB Scheme with the U-PACE electronics (certificate CA38941CSA). Transmitter mounting code 10C in the order details designates the version supplied from the factory with both approvals integrated.
Wetted components are the epoxy-coated ASTM A105 steel process connection (corrosivity category C4M), an NBR liner with other liner materials available on request, and Hastelloy C276 measuring electrodes. The electrodes use galvanic tapping and sit in direct contact with the medium.
Yes. Magnetic flow meter, electromagnetic flow meter, and magnetic inductive flowmeter are three names for one technology, and KOBOLD documentation uses all three for the MIS. Each describes an instrument applying Faraday's law of induction to conductive liquids, so specifications stay identical whichever search term brought you here.
Product Specifications
Measuring Range
2...277 GPM to 18...4,400 GPM
Max. Media Viscosity
70 cP
Process Connection
2", 3", 4", 6" or 8" ANSI Flange
Maximum Pressure
232 PSIG
Repeatability
± 0.2% of Full Scale
Accuracy
<± (0.5% of reading + 0.5% of full scale)
Product Features
Monitoring, Batching, and Transmitter Functions
Grand and Resettable Totalizer
Batching Function has an External Control Input
Bi-directional Measurement
Documents
Software Downloads
There are no downloads to display at this time. If you think this is incorrect, please contact us and let us know the issue you are having.
Videos
C3T0 Electronics Technology Explained
Join KOBOLD Applications Engineer Michael Besser for a comprehensive look at C3T0 technology and its practical applications in flow measurement. In this informational session, we'll explore how C3T0 technology works and where it's implemented across KOBOLD's product line. There are two configurable outputs and a grand and resettable totalizer. Temperature measurement is also integral and bi-directional flow measurement is also possible. To learn more about our U-PACE C3T0 electronics package,
view our U-PACE Video.