The KOBOLD KFF/KFG is a low volume turbine flow meter that measures very low flow rates of liquids (KFF) and gases (KFG) with a Pelton-type turbine wheel and photodiodes. It is designed for industrial, commercial and laboratory flow applications involving media transparent to the near-infrared spectrum, and it provides a linear 0-5 VDC analog output, with a pulse output available as an option.
Two series are offered. Series 1000 flow sensors deliver the analog signal through the included cable and are designed for use with data acquisition equipment. Series 3000 flowmeters add a compact 0.4" (10.2 mm) in-line LCD display. Both series are available in Ryton® (PPS) or brass construction.
How the KFF/KFG Flow Meter Works
The KFF/KFG operates on the Pelton principle. The flowing medium causes a vane to rotate, and photodiodes convert that rotary motion into electrical impulses. The electronics then output the flow rate as a linear analog signal or, with Option T, as a frequency signal with a maximum of 500 Hz. The medium must be transmissive to infrared light.
KFF for Liquids and KFG for Gases
KFF models are specified for water and cover liquid flow ranges from 13 to 100 ml/min up to 1 to 10 l/min. KFG models are calibrated in air at standard conditions (14.7 psia and 68 °F; 1.013 bar abs and 20 °C) and cover gas flow ranges from 20 to 100 ml/min up to 100 to 500 l/min.
KFF units are intended for low-viscosity liquids below 10 mm²/s (10 cSt); liquids with higher viscosity can produce significant measurement errors. The media may not be contaminated, since large fiber particles can jam or destroy the rotor.
Series 1000 Flow Sensors vs. Series 3000 Flowmeters
Series 1000 flow sensors provide a 0-5 VDC output, operate from 11.5 to 15 VDC at 50 mA typical, and are available with Option T for a combined 0-5 VDC and pulse output. Series 3000 flowmeters add a 3 1/2 digit LCD display with 0.4" (10.2 mm) characters and need just 35 mA at 12.5 VDC ± 2 VDC.
Because of the low current draw, a Series 3000 flowmeter can run from a battery. Option T is offered on KFF models only and cannot be combined with Option B2 (18.0 to 25.0 VDC power).
Flow Ranges, Fittings and Pressure Drop
KFF liquid sensors come in seven ranges with 1/8", 1/4" or 3/8" compression tube fittings, and KFG gas sensors come in twelve ranges with 1/8" through 1/2" compression tube fittings. Pressure drop varies by model: 6 to 15 psi (0.41 to 1.03 bar) across the KFF range and 2 to 20 inches of water column (5.0 to 49.8 mbar) across the KFG range.
A brass body is offered on all KFF models and on KFG 3000 flowmeters through the 2 to 10 l/min range; KFG 1000 sensors are Ryton only.
Accuracy, Pressure and Temperature Ratings
Both accuracy and linearity are rated at ±3% of full scale. Repeatability for liquids is ±0.2% of full scale on both series; for gases it is ±1% of full scale (Series 1000) and ±0.5% of full scale (Series 3000). The operating temperature range is 32 to 120 °F (0 to 50 °C) with a drift of ±0.1% per °F (±0.18% per °C).
Maximum pressure depends on material and media. KFF liquid models: 100 PSIG (6.9 bar) with a Ryton body and 500 PSIG (34.5 bar) with a brass body. KFG gas models: 40 PSIG (2.8 bar) in all materials.
Wetted Materials
Ryton units use a housing and turbine of 60% Ryton / 40% glass, a sapphire bearing, a 316 stainless steel support, Viton® (FKM) gaskets and acetal fittings. Brass units keep the same turbine, bearing, support and gaskets but use a brass and glass housing and brass fittings. EPDM O-rings (Option OR3) and brass (Option C1) or stainless steel (Option C2) compression fittings are available.
Installation and Wiring
The KFF/KFG is installed horizontally. KFG gas models are calibrated in the horizontal position with the type label facing upward, and other mounting positions can introduce a measurement deviation. Use two suitable wrenches when making connections so that no torque is transferred to the housing.
Before start-up, confirm that the actual flow is within the measuring range on the type label; continuous operation outside the range can damage the bearings. Large air bubbles can cause error indications and bearing damage, so the sensor must stay completely filled with media. Let media enter slowly, and protect the unit from water hammer caused by solenoid (magnetic) valves or ball valves.
Wiring is made with the supplied connection cable, and the plug and socket are coded so that a wrong connection is not possible. Incorrect wiring will damage the electronics: confirm the supply voltage matches the unit and de-energize the supply wires before connecting.
Common Questions about KFF/KFG - Low Volume, Rotating Vane Flow Meters
KFF models measure liquids and KFG models measure gases. KFF ranges start at 13 to 100 ml/min and end at 1 to 10 l/min. KFG ranges start at 20 to 100 ml/min and end at 100 to 500 l/min. KFG units are calibrated in air, and only the KFF is available with the Option T pulse output.
The standard output is a 0-5 VDC analog signal that is linear with flow rate and accessible via the included cable. Option T adds a pulse output with a maximum frequency of 500 Hz on KFF liquid models. Option T cannot be combined with the Option B2 18.0 to 25.0 VDC power supply.
Series 1000 sensors run on 11.5 to 15 VDC and draw 50 mA typical. Series 3000 flowmeters run on 12.5 VDC ± 2 VDC and draw 35 mA typical. Option B2 provides an 18.0 to 25.0 VDC supply instead, and 110 VAC (KF-8000) or 230 VAC (KF-8100) adapters are offered for 12 VDC models.
Liquid KFF models are rated to 100 PSIG (6.9 bar) in Ryton and 500 PSIG (34.5 bar) in brass. Gas KFG models are rated to 40 PSIG (2.8 bar) regardless of material. All models operate between 32 and 120 °F (0 and 50 °C), and the maximum operating pressure and temperature must not be exceeded.
Accuracy is ±3% of full scale, and linearity is the same. Liquid repeatability is ±0.2% of full scale on either series. Gas repeatability is ±1% of full scale with Series 1000 sensors and ±0.5% with Series 3000 flowmeters. Temperature drift is ±0.1% for each °F (±0.18% for each °C), and the sensors have no zero drift.
Infrared light must be able to pass through the medium. KFF liquid models are limited to viscosities under 10 mm²/s (10 cSt), because higher viscosity causes significant measurement errors. The media must be free of contamination: large fiber particles can cause jamming or destruction of the rotor.
Mount the meter in a horizontal position. The KFG gas versions are calibrated with the type label facing up, and a different orientation can affect the reading. Series 1000 sensors include two mounting flanges. Tighten connections with two wrenches, keep the unit completely filled with media, and avoid air bubbles and pressure peaks from fast-acting valves.
Housings are either 60% Ryton / 40% glass (PPS) or brass with glass. In both versions the turbine is 60% Ryton / 40% glass, the bearing is sapphire, the support is 316 stainless steel and the gaskets are Viton® (FKM). Fittings are acetal on Ryton units and brass on brass units; EPDM O-rings are optional (OR3).
Yes. The Series 3000 flowmeter requires only 35 mA at 12 VDC, so it can be battery powered or connected to a 12 volt source with the included cable. Its 3 1/2 digit LCD has 0.4" (10.2 mm) characters, and the unit keeps the same 0-5 VDC analog output as the Series 1000.
Pressure drop depends on the model. KFF Series 1000 liquid sensors drop 7 to 15 psi (0.48 to 1.03 bar), and KFF Series 3000 flowmeters drop 6 to 10 psi (0.41 to 0.69 bar). KFG gas models drop between 2 and 20 inches H2O (between 5.0 and 49.8 mbar), with most gas ranges at 2 or 3 inches H2O.
Product Specifications
Measuring Range
13...100 ml/min to 1...10 l/min (liquid) / 20...100 ml/min to 100...500 l/min (gas)
Fittings
1/8"...1/2" Tube Fittings
Wetted Materials
PPS, Brass, FKM
Max. Pressure
500 PSIG (Brass KFF)
Max. Temperature
120 °F (50 °C)
Media Categories
KFF for Liquid, KFG for Gas
Product Features
Series 1000 Flow Sensors - 0–5 VDC Output - Highly Repeatable - 12.5 VDC Input Power
Series 3000 Flowmeters - All Features of the 1000 Series Plus Local LCD Display
Documents
Software Downloads
There are no downloads to display at this time. If you think this is incorrect, please contact us and let us know the issue you are having.
Videos
There are no videos to display at this time. If you think this is incorrect, please contact us and let us know the issue you are having.